6. Commissioning, operation
SWITCH TEST mode:
SWITCH TEST can be performed.
With [▲] or [▼] the function can be
switched between pressure signal and
EN
electrical signal.
Current measured value of the
1
CPT6000 reference pressure
sensor
Current switching status/status of
2
the pressure switch
Opening switch point
3
Closing switch point
4
Hysteresis/separation between
5
opening and closing of the switch
Pressure unit (adjustable via menu)
6
6.9

Additional setup menu

6.9.1 Functions

With the [▲] or [▼] button, select menu item.
Input via numeric keypad
Confirm by pressing [ENTER].
[CLEAR] deletes the input, or resets the Min./Max. memory
Ref.:
1
Current measured value of the connected CPT6000 reference pressure
sensor
Tare:
2
Offset function that adjusts the current measured value. The value
entered is added to the current measured value.
(e.g. Ref. 0.000 and Tare: 1.000 → [new] Ref. 1.000)
Min./Max.:
3
Minimum and maximum value memory
The memory is reset by highlighting the value with the [▲] or [▼] cursor
and pressing [CLEAR].
Alarm:
4
Audible and visible alarm function
Upper alarm limit: ≥ bar
Lower alarm limit: ≤ bar
If the current measured value goes outside the set alarm limits, an intermittent signal tone sounds and the lower status
line blinks.
Activation
Move the cursor to the OFF field, next to the word Alarm, and via [◀] or [▶] change it to ON.
Deactivation
Setting it back to OFF
Filter [1-5]:
5
Damping/smoothing of the reference sensor signal
Definition of the numbers:
1 = no additional smoothing ... 5 = strong smoothing
38
Switch-Test
13:54
Reference
0.000
10.000
1
0.000
bar
2
Status:
3
0.000
bar
0.000
bar
4
5
Hys:
0.000
bar
Before the pressure
switch test
WIKA operating instructions, pressure calibrator model CPH6000
Switch-Test
13:54
Reference
0.000
10.000
4.200
bar
Status:
4.900
bar
5.000
bar
Hys:
0.100
bar
After the pressure switch
test
..Functions
bar
Ref.:
0.000
1
Tare:
0.000
bar
2
Min:
0.000
bar
3
Max:
0.000
bar
Alarm:
ON
4
10.00
bar
bar
-1.00
0
Filter:
5
Ref. = rel. Pressure
Offset pressure