6. Commissioning, operation
SWITCH TEST mode:
▶
SWITCH TEST can be performed.
▶
With [▲] or [▼] the function can be
switched between pressure signal and
EN
electrical signal.
Current measured value of the
1
CPT6000 reference pressure
sensor
Current switching status/status of
2
the pressure switch
Opening switch point
3
Closing switch point
4
Hysteresis/separation between
5
opening and closing of the switch
Pressure unit (adjustable via menu)
6
6.9
Additional setup menu
6.9.1 Functions
▶
With the [▲] or [▼] button, select menu item.
▶
Input via numeric keypad
▶
Confirm by pressing [ENTER].
▶
[CLEAR] deletes the input, or resets the Min./Max. memory
Ref.:
1
Current measured value of the connected CPT6000 reference pressure
sensor
Tare:
2
Offset function that adjusts the current measured value. The value
entered is added to the current measured value.
(e.g. Ref. 0.000 and Tare: 1.000 → [new] Ref. 1.000)
Min./Max.:
3
Minimum and maximum value memory
The memory is reset by highlighting the value with the [▲] or [▼] cursor
and pressing [CLEAR].
Alarm:
4
Audible and visible alarm function
Upper alarm limit: ≥ bar
Lower alarm limit: ≤ bar
If the current measured value goes outside the set alarm limits, an intermittent signal tone sounds and the lower status
line blinks.
Activation
■
Move the cursor to the OFF field, next to the word Alarm, and via [◀] or [▶] change it to ON.
Deactivation
■
Setting it back to OFF
Filter [1-5]:
5
Damping/smoothing of the reference sensor signal
Definition of the numbers:
1 = no additional smoothing ... 5 = strong smoothing
38
Switch-Test
13:54
Reference
0.000
10.000
1
0.000
bar
2
Status:
3
0.000
bar
0.000
bar
4
5
Hys:
0.000
bar
Before the pressure
switch test
WIKA operating instructions, pressure calibrator model CPH6000
Switch-Test
13:54
Reference
0.000
10.000
4.200
bar
Status:
4.900
bar
5.000
bar
Hys:
0.100
bar
After the pressure switch
test
..Functions
bar
Ref.:
0.000
1
Tare:
0.000
bar
2
Min:
0.000
bar
3
Max:
0.000
bar
Alarm:
ON
4
10.00
bar
bar
-1.00
0
Filter:
5
Ref. = rel. Pressure
Offset pressure
6