WIKA CPH6000 Betriebsanleitung Seite 29

Prozesskalibrator
Vorschau ausblenden Andere Handbücher für CPH6000:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

ProcessCalibrator
CPH6000
d) Functions: Operating functions such as:
Tare: Offset correction for reference pressure sensors
Min/Max: Minimum/Maximum memory
Alarm: Min/Max alarm (visual & audible)
Filter: Smoothing the signal of the reference sensor
For further information see Chapter 10.1
e) CPH-Info: General CPH6000 instrument data such as:
Calibration data for the electrical measurement inputs
Firmware number
Serial-No. of the device
For further information see Chapter 10.2
f) Reference Sensor: Data for the currently connected reference pressure
sensor such as:
Measurement range
Accuracy class
Sensor measurement type
Information in the event of reference sensor overpressure
Calibration data for the reference sensor
For further information see Chapter 10.3
g) Reference Sensor list:
List of the stored reference sensors that can be attached and calibrated.
For further information see Chapter 10.4
h) CPH Settings:
Info: on battery level
Display options of: Menu language, system time / system clock, display-
brightness, powersave function (automatic energy saving mode; see
Chapter 10.5)
Entry/Input options:
- ambient temperature during the calibration (see Chapter 5.2)
- height difference between reference and pressure sensor
(see Chapter 5.2).
For further information see Chapter 10.5
i) Digital Interface:
 Switch between USB and RS-232 digital interfaces incl. baud rate entry
For further information see Chapter 10.6
j) Clear CalData:
 All stored calibration data will be deleted (Clear complete storage space)
For further information see Chapter 10.7
WIKA Operating Instructions ProcessCalibrator
GB
29