Type of fault
V. A intermittent audio alarm is audible
and the display only shows lines
rather than the measured value.
VI. In the lower information line of the
display is showing the message
"low BAT".
VII. The backlight and the internal 24 V
test-item power supply are suddenly
no longer working.
VIII. The test specimen (read through
the mA and/or V-measuring input)
shows an incorrect value and/or
doesn't respond.
Further help can be found through WIKA's Test & Calibration Technology Depart-
ment
WIKA Operating Instructions ProcessCalibrator
Measures
The current pressure is more than
10 % over or below the allowable pres-
sure range. Adjust the pressure immedi-
ately and appropriately.
The battery charge level is less than
10 %. The instrument must be charged
using its charger unit.
Press any button (other than ON/OFF)
in order to cancel the Power-Save
function, and if needs be, increase the
Countdown-time of the Power-save
function. (see Chapter 10.5)
Examine the wiring; examine whether
the test specimen's measuring range
was entered correctly or, with a test
specimen without its own power
supply, whether the 24 V test specimen
supply voltage is turned on.