Herunterladen Inhalt Inhalt Diese Seite drucken

HBM U9C Montageanleitung Seite 12

Kraftsensoren
Vorschau ausblenden Andere Handbücher für U9C:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Scope of supply, accessories and equipment variants
2. Device plug/electrical connection
We can mount one of the following connectors on the U9C if requested:
- D-SUB connector, 15-pin: 15-pin connector for connection to numerous
amplifier systems, e.g. MGCplus, Scout, MP85, etc.
- D-SUB HD connector: A 15-pin plug for connection to appropriate
amplifier systems, e.g. the HBM QuantumX system
- 3106 PEMV connector (Greenline): For connection to appropriate
amplifier systems, e.g. MGCplus with AP03.
- Free ends: Transducer delivered without plugs.
- The force transducer can be ordered with a permanently connected
amplifier electronics module which converts the output signal of the force
transducer into a 0 ... 10 V signal. This option is only available with 1.5 m
and 3 m cable lengths.
- The force transducer can also be ordered with a permanently connected
amplifier electronics module which converts the output signal of the force
transducer into a 4 ... 20 mA signal. This option is only available with
1.5 m and 3 m cable lengths.
3. TEDS
You can order the force transducer with transducer identification ("TEDS").
TEDS (Transducer Electronic Data Sheet) allows you to store the
transducer data (characteristic values) in a chip that can be read out by a
connected measuring device (with an appropriate amplifier). HBM records
the TEDS data before delivery so that no parameterization of the amplifier
is necessary.
TEDS can only be fitted in the plug of the U9C, therefore it is not possible to
equip the "free cable ends" version with TEDS. The versions with
permanently connected amplifier electronics cannot be connected with the
TEDS option.
10
A03815_03_YI0_01 HBM: public
U9C

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis