Herunterladen Inhalt Inhalt Diese Seite drucken

Information On The Power Supply/Battery Charging - Ottobock 1B1-2 Gebrauchsanweisung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 1
If the product and its components have been subjected to extreme loads (e.g. due to a fall,
etc.), then the product must be inspected for damage immediately by a prosthetist. If neces­
sary, the prosthetist will forward the product to an authorised Ottobock Service Centre.
CAUTION
Signs of wear and tear on the product components
Falling due to damage or malfunction of the product.
Regular service inspections (maintenance) are mandatory in the interest of your own safety
and in order to maintain operating reliability and protect the warranty.
NOTICE
Improper product care
Damage to the product due to the use of incorrect cleaning agents.
Only clean the product with a damp cloth and mild soap (e.g. 453H10=1 Ottobock Der­
maClean).
NOTICE
Mechanical damage to the product
Change in or loss of functionality due to damage.
Use caution when working with the product.
If the product is damaged, check it for proper function and readiness for use.
In case of changes in or loss of functionality, do not continue using the product (see "Signs
of changes in or loss of functionality during use" in this section).
Take any necessary measures (e.g. repair, replacement, inspection by the manufacturer's
customer service, etc.).
CAUTION
Use of unapproved accessories
>
Falling due to product malfunction as a result of reduced interference resistance.
>
Interference of other electronic devices due to increased emissions.
Use the product only in combination with the accessories, signal converters and cables lis­
ted in the sections "Scope of delivery" (see Page 57) and "Accessories" (see Page 57).
INFORMATION
When exoprosthetic components are used, the use of control functions which are performed
hydraulically or movement of the component in the footshell can cause noise. This kind of noise
is normal and unavoidable. It generally does not cause any problems. If movement noise
increases noticeably during the lifecycle of the component, the component should be immedi­
ately inspected by authorised qualified Ottobock personnel.
Signs of changes in or loss of functionality during use
Decreased forefoot resistance or changes in roll-over behaviour are noticeable indications of loss
of functionality.

4.4 Information on the Power Supply/Battery Charging

CAUTION
Charging the prosthesis without taking it off
>
Falling due to walking and getting caught on a connected battery charger.
52

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis