OBJ_DOKU-184600-004.fm Seite 21 Montag, 14. September 2015 11:18 11
IMPORTANT! If you have called up a different program to the ba-
sic program when you inserted the handle, the instrument de-
tection facility on the handle will set the modes and values for
effect and performance according to the program you have se-
lected.
4.6
ReMode button
The ReMode button is used for switching between programs
with two setting options (e.g. 1a and 1b), each of which permits
individually configurable setting of the operating parameters,
Mode, Effect and Rated power. Both settings can also be given a
name.
5
Cleaning, disinfection, sterilization
5.1
Safety instructions
May be cleaned in an ultrasonic bath.
Never use sharp objects for cleaning.
Disinfectants must be rinsed off thoroughly after use.
Maximum air pressure during drying: 2 bar.
Maximum water pressure during rinsing: 2 bar.
During machine cleaning and disinfection, do not exceed 95 °C.
It is essential to follow the cleaning agent and disinfectant man-
ufacturers' instructions.
During sterilization, do not exceed 138 °C.
Do not sterilize in hot air.
21