Herunterladen Inhalt Inhalt Diese Seite drucken

Cleaning, Disinfection, Sterilization; Safety Instructions; Reprocessing Limitation; Recommended Equipment / Substances - ERBE 20195-080 Verwendungshinweis

Bipolare zangen
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 1
OBJ_DOKU-25953-005.fm Seite 20 Dienstag, 28. Juli 2015 1:13 13
5

Cleaning, disinfection, sterilization

5.1

Safety instructions

Erbe Elektromedizin recommends machine cleaning/
disinfection. Because it is much less effective, manual cleaning
is not recommended.
On no account use sharp or abrasive objects for cleaning purpos-
es.
Disinfectants must be rinsed off thoroughly after use!
The product can be cleaned/disinfected up to a maximum tem-
perature of 95 °C.
It is essential to follow the cleaning agent and disinfectant man-
ufacturers' instructions.
Can be sterilized up to a maximum temperature of 134 °C.
Do not sterilize in hot air!
5.2

Reprocessing limitation

Frequent reprocessing can damage this product. Observe the
safety instructions for checking the product. With visible damage
or functional impairment, the product may no longer be used.
5.3

Recommended equipment / substances

The service life of the product, cleaning action and sterility are
validated for the following cleaning and preparation procedures
if the recommended or equivalent substances are used in accor-
dance with the manufacturer's instructions.
• Manual cleaning/disinfection with the detergent
Cidezyme ® LF/Enzol and disinfectant Cidex ® OPA (John-
son & Johnson Medical Limited, Gargrave, Skipton).
20

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

20195-08620195-09220195-09320195-081

Inhaltsverzeichnis