Agilent Technologies 6890N Benutzerhandbuch Seite 283

Gaschromatograph
Inhaltsverzeichnis

Werbung

Warnhinweise
Tabelle 26 (Fortsetzung)
Warncode
Statusmeldung
111
Col 1 config changed
112
Col 2 config changed
113
Aux 3 method changed
114
Aux 4 method changed
115
Aux 5 method changed
116
117
F inl calib deleted
118
B inl calib deleted
119
F det calib deleted
120
B det calib deleted
121
P aux calib deleted
122
Comm data overrun
123
Comm data error
124
Comm abnormal break
Popup-Meldung
Column 1: config changed,
method defaulted
Column 2: config changed,
method defaulted
Aux 3 config changed
Method defaulted
Aux 4 config changed
Method defaulted
Aux 5 config changed
Method defaulted
F inl calib deleted
B inl calib deleted
F det calib deleted
B det calib deleted
P aux calib deleted
Host communications:
data overrun
Host communications:
data error
Host communications:
abnormal break
Eintrag im Logbuch
Log overflow
Possible data loss:
Comm data overrun
Possible data loss:
Comm data error
Possible data loss:
Comm abnormal break
Meldungen
Bemerkungen
Korrigieren Sie die
Methode so, dass Sie der
Hardware entspricht.
Korrigieren Sie die
Methode so, dass Sie der
Hardware entspricht.
Korrigieren Sie die
Methode so, dass Sie der
Hardware entspricht.
Korrigieren Sie die
Methode so, dass Sie der
Hardware entspricht.
Korrigieren Sie die
Methode so, dass Sie der
Hardware entspricht.
Die Logbuch-Kapazität
liegt bei 50 Einträgen.
Der Einlass wechselt
wieder zur
Standardkalibrierung.
Der Detektor wechselt
wieder zur
Standardkalibrierung.
Das Aux-Modul wechselt
wieder zur
Standardkalibrierung.
Wenden Sie sich an den
Agilent Kundendienst.
Wenden Sie sich an den
Agilent Kundendienst.
Überprüfen Sie die
Kabelanschlüsse.