Expected Service Life - Stryker 1588-020-122 Handbuch

Aim coupler, 18 mm, c-mount
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 33
Steris System 1/1E /1 Plus /1 Express
Note: Steris System 1, System 1 Plus, and System 1 Express are
not intended for use in the United States.
1. Clean and prepare the coupler as recommended in this user
manual.
2. Following the instructions of the manufacturer, sterilize
the coupler using one of the Steris systems below with the
appropriate sterilant:
• System 1 with Steris 20 Sterilant
• System 1E with S40™ Sterilant
• System 1 Plus with S40 Sterilant
• System 1 Express with S40 Sterilant
3. Remove the coupler from the Steris chamber once sterilization
is complete, or moisture may condense inside the coupler and
cause fogging.
4. Allow the coupler to completely dry and cool to room
temperature before reassembling it to a scope or camera
head. Otherwise, the lens can fog during use.
Maintenance
Inspection
• Inspect the device on a continual basis. If a problem is observed
or suspected, the device should be returned for evaluation and/or
repair.
• Inspect all components for cleanliness. If fluid or tissue buildup is
present, repeat the above cleaning and sterilization procedures.
Storage
Store the coupler in a dry, clean, and dust-free environment at room
temperatures.

Expected Service Life

When the device is sterilized with V-PRO, Sterrad, or Steris
System 1/1E/1 Plus/1 Express, the expected service life is
280 reprocessing cycles (two years based on one cycle per case,
and 140 cases per year).
When the device is sterilized with Ethylene Oxide, the expected
service life is 140 reprocessing cycles (two years based on one cycle
per case, and 70 cases per year).
When the device has reached the end of its service life, dispose of it
according to local laws and hospital practices.
EN-11