Disposable
RF probe is not
ready
Attachments
RF probe is expired
RF probe
identification is
invalid
RF probe
communication
error
Exceeded time
usage
RF power is too
high
RF voltage is too
high
RF current is too
high
RF delivery
has exceeded
continuous limit
Low impedance
detected
Footswitch
A wireless
footswitch is
detected
The footswitch icon
does not appear
Note: If a disturbance occurs on the video monitor, the user should ensure that
the probe cable is not near any other instrument cables.
EN-34
Check the connection to the
console.
Replace probe.
Replace probe.
• Check the connection to the
console.
• If necessary, replace probe.
Replace probe
• Check the probe for damage.
• If necessary, replace probe.
• Check the probe for damage.
• If necessary, replace probe.
• Check the probe for damage.
• If necessary, replace probe.
Clear error and continue
• Check the probe for damage.
• If necessary, replace probe.
Disconnect the wired footswitch.
• Ensure the unit is connected.
• Ensure that there is no damage to
the cable or connector.