Guidance and Manufacturer's Declaration: Electromagnetic Immunity
Crossfire™ System is intended for use in the electromagnetic environment specified below. The
customer or the user of Crossfire™ System should ensure that it is used in such an environment.
Immunity Test
Conducted RF
3 Vrms 150 kHz to
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
80MHz to 2.5 GHz
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
(a) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast, and TV broadcast,
cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due
to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured
field strength in the location in which the Crossfire™ System is used exceeds the applicable RF
compliance level above, the Crossfire™ System should be observed to verify normal operation. If
abnormal performance is observed, additional measures may be necessary, such as reorienting or
(b) Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
EN-14
IEC 60601 Test
Compliance Level
Level
80 MHz
3 V/m
absorption and reflection from structures, objects, and people.
relocating the Crossfire™ System.
Electromagnetic Environment:
communications equipment should
be used no closer to any part of
the Crossfire™ system, including
its cables, than the recommended
separation distance calculated
from the equation applicable to the
Recommended Separation Distance
3 V
3 V/m
where P is the maximum output
power rating of the transmitter
in watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance in
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey (a), should
be less than the compliance level in
Interference may occur in the vicinity
of equipment marked with the
Guidance
Portable and mobile RF
frequency of the transmitter.
d = 1.17 √P
d = 1.17 √P
80 MHz to 800 MHz
d = 2.33 √P
80 MHz to 2.5 GHz
meters (m).
each frequency range(b).
following symbol: