Recommended Separation Distances Between Portable and Mobile RF Communications
The Crossfire™ System is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The user of the Crossfire™ System can help prevent electromagnetic
interference by maintaining a minimum distance between portable and mobile RF communications
equipment (transmitters) and the Crossfire™ System as recommended below, according to the
maximum output power of the communications equipment.
Rated maximum
output power (W) of
transmitter
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance (d)
in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
Equipment and the Crossfire™ System
Separation distance (m) according to frequency of transmitter
150 kHz to 80 MHz
0.12
0.37
1.17
3.70
11.70
reflection from structures, objects, and people.
80 MHz to 800 MHz
0.12
0.37
1.17
2.33
11.70
800 MHz to 2.5 GHz
0.23
0.74
3.70
7.37
23.30