12.3 ELECTROMAGNETIC IMMUNITY / RADIOFREQUENCY MOBILE EQUIPMENT

The device is intended for use in the electromagnetic environment specified in the table below. The user and/or
installer must ensure that the device is used in such an electromagnetic environment.
Immunity test
Portable and mobile radiofrequency communications devices must not be used near the device (including its
cables) at a distance less than that recommended and calculated according to the frequency and power of the
Conducted disturbance,
radiofrequency fields.
IEC61000-4-6
Radiated radiofrequency
electromagnetic field.
IEC61000-4-3
The electromagnetic field strengths of fixed radiofrequency emitters, as determined by an electromagnetic
environment measurement (a), must be less than the compliance level in each frequency range (b).
Interference may occur near equipment marked with the symbol below:
Note 1: At 80MHz and 800MHz, the higher frequency range applies.
Note 2: These specifications may not be applicable in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and persons.
(a): The electromagnetic field strengths of fixed radiofrequency emitters, such as base stations for mobile
telephones (cellular/cordless), mobile radios, amateur radio, AM/FM radio broadcasts and TV broadcasts cannot
be determined exactly by theory. To assess the electromagnetic environment due to fixed radiofrequency
emitters, an electromagnetic environment measurement must be made. If the measured radiofrequency field
strength in the immediate environment where the product is used exceeds the compliance level specified above,
the performance of the product must be tested to verify whether it conforms to the specification. If abnormal
performance is observed, additional measures may be necessary, such as reorienting or relocating the product.
(b): In the 150 kHz to 80 MHz frequency range, the electromagnetic field strengths must be less than 3 V/m.
IEC 60601
Compliance
test level
level
3 V/m
150kHz to
3 V/m
80MHz
3 V/m
80MHz to
3 V/m
2.5GHz
Electromagnetic environment - guidance
emitter.
Recommended separation distance:
d = 1.2
 P
d = 1.2
80MHz to 800MHz.
 P
d = 2.3
800MHz to 2.5GHz.
 P
Where P is the maximum power rating of the emitter
in watts (W) according to the manufacturer's
specifications and d is the recommended minimum
separation distance in meters (m).