11.2 PREVENTIVE AND CORRECTIVE MAINTENANCE

Micromotors inspection/overhaul: The manufacturer recommends inspection or overhaul of the micromotors at
least once a year.
The micromotor cable has limited durability and is considered a consumable product. The durability of the cable
is largely dependent upon use and treatment, i.e., handling, reprocessing and frequency of use. Caution must be
exercised when connecting or disconnecting the micromotor and foot control cable from the device to avoid
damage to the cable (Fig. 8). Motor assembly has been tested for 500 sterilization cycles.
Periodically inspect the micromotor cord for damage prior to use. If there is damage noted, contact SATELEC
1. Do not attempt to disassemble the micromotor or its connector.
2. Do not oil or lubricate the micromotor.
3. Do not attach a handpiece to the micromotor while the motor is running.
4. Do not bend micromotor cord sharply.
5. Do not drop or impact micromotor against hard surface.
Failure to comply with any of the above instructions may avoid your warranty
The average life of this medical device with normal use and in accord with the present user
manual, has been calculated to be 5 years. According with EN62353 we recommend users verify
every 2 years from the date of the first use: dispersion current versus ground, in the box and in
the patient.
MOTOR RUBBER O-RING:
 If necessary, substitute the "Rubber O-Ring"
LUBRICATING:
 Do not oil or lubricate the micromotor. Protect motor from excess oil draining from handpiece. After
lubricating and before autoclaving, stand handpiece by its base on a paper towel and allow excess oil to
drain.

11.3 FUSE REPLACEMENT

Your device is protected by one fuse. The replacement procedure is as follows:
ONCE
UNIT IS UNPLUGGED,
-
remove
the fuse holder.
-
- Extract
the
fuse holder, replace the fuse with one of the same type and value 1x4A T (Dim. 5x20).
Note: All the information necessary for the repair of faulty components is made available by the manufacturer to
the technical personnel of the
 The I- S
URGE
 The I- S
URGE
repairs must only be undertaken by an authorized service organization.
 Unauthorized
Use on
electrical
®
SATELEC
approved dealer network who carry out the work.
must
be used in accordance with these Instructions for Use.
has
no components which can be repaired by the user. Assembly, modifications or
opening
of the equipment invalidates all claims under warranty and any other claims.
circuit complying with the terms of regulation CEI 64-8 section 710.
®
.
18