Fault detected: The display is not working properly
Possible cause
Contra-angle damaged.
Wrong voltage.
Possible interference.
Broken connections inside.
Display damaged.
PCB damaged.
Fault detected: The keyboard is not working properly
Possible cause
Keyboard damaged.
Broken connections inside.
Fault detected: The sound signals are not working properly
Possible cause
Broken connections inside
Fault detected: The motor doesn't work
Possible cause
Micromotor
cable not
connected properly
Contra-angle defective or
not properly calibrated.
Footswitch damaged.
Wrong voltage.
PCB damaged.
Motor damaged.
Connector pins and/or
cable damaged.
Fault detected: The motor loses power.
Possible cause
Contra-angle defective or
not properly calibrated.
Motor damaged.
Fault detected: The motor stops suddenly
Possible cause
Micromotor/Cable
damaged
Wrong voltage.
Footswitch damaged.
PCB damaged.
Solution
Check the integrity of the Contra-angle.
Verify that the voltage matches the power supply. If it does, try to switch the central unit off
and then on again, or connect the device to another electrical outlet.
Check to see if the central unit has been positioned too close to another electrical device
which can cause interference. In this case, move the interfering electrical device further away
from the central unit.
If you cannot find the cause of the malfunction, immediately discontinue use of the device and
®
SATELEC
contact
after-sales service.
Solution
Attempt to change parameters by using the footswitch.
®
SATELEC
Contact
after-sales service.
Solution
Immediately discontinue use of the device and contact
Solution
Check to see that the micromotor cable is plugged into the connector on the front right of the
central unit, disconnect and reconnect.
Check the integrity of the Contra-angle. Remove the Contra-angle and perform calibration.
micromotor
Attempt to start the
keyboard. If unit does not turn on, contact
Verify that the voltage matches the power supply. If it does, try to switch the central unit off
and then on again, or connect the device to another electrical outlet.
Immediately discontinue use of the device and contact
Solution
Check the integrity of the Contra-angle by removing the Contra-angle and perform calibration.
®
SATELEC
Contact
after-sales service.
Solution
Check to see that the micromotor cable is plugged into the connector on the front right of the
central unit, disconnect and reconnect. Start the motor by pressing the [Motor ON/OF] button
on the keyboard. If the motor turn on, immediately discontinue use of the device and contact
®
SATELEC
after-sales service.
Verify that the voltage matches the power supply. If it does, try to switch the central unit off
and then on again, or connect the device to another electrical outlet.
Attempt to start the Micromotor by using the [Motor ON/OFF button] on the central unit
keyboard.
If unit does not turn on, contact
If you cannot find the cause of the malfunction, immediately discontinue use of the device and
contact dealer, the authorized service center, or the manufacturer directly.
SATELEC
by using the [Motor ON/OFF button] on the central unit
®
SATELEC
after-sales service.
SATELEC
®
SATELEC
after-sales service.
®
after-sales service.
®
after-sales service.
20