Start of QS Volume Wizard
Shows the type of Wizard to be executed.
Options
Transducer
Specifies the Siemens transducer connected to the device.
Options
Temperature Source
Source of the temperature reading used to adjust the speed of sound.
Options
Fixed Temperature
Use this feature if a temperature sensing device is not used.
Value
This parameter is only displayed if FIXED TEMPERATURE selected for
Source
.
Vessel Shape
Defines the vessel shape and allows the SITRANS LUT400 to calculate volume
instead of level. If NONE is selected, no volume conversion is performed. Select the
vessel shape matching the monitored vessel or reservoir.
Options
If CURVE TABLE or LINEAR TABLE selected, enter values for level and volume
breakpoints after completing the wizard. (See parameter
Operating Instructions.)
Units
Sensor measurement units.
Options
For the purpose of this example, all values are assumed to be in meters (m).
Note:
7ML19985XU81
CANCEL, START
NO TRANSDUCER, XRS-5, XPS-10, XPS-15, XCT-8, XCT-12, XPS-30, XPS-40,
XLT-30, XLT-60, STH
Default: NO TRANSDUCER
TRANSDUCER, FIXED TEMPERATURE, EXTERNAL TS-3, AVGERAGE OF
SENSORS
Default: TRANSDUCER
Range: -100.0 to +150.0 °C
Default: +20.0 °C
NONE, LINEAR, CYLINDER, PARABOLIC BOTTOM, HALF SPHERE BOTTOM,
FLAT SLOPED BOTTOM, PARABOLIC ENDS, SPHERE, CONICAL BOTTOM,
CURVE TABLE, LINEAR TABLE
Default: LINEAR
M, CM, MM, FT, IN
Default: M
SITRANS LUT400 (HART)
Temperature
2.6.7.Table 1-8
in full
Page EN-21