Wiring compartment
The terminal board on the LUT400 allows all inputs and outputs to be connected
simultaneously. Terminal strips can be removed to improve ease of wiring.
CAUTION: Ensure the terminal strips are terminated to the correct location
during re-installation. Failure to do so may result in damage to the device, or
external equipment that is attached.
Note: Recommended torque on terminal clamping screws.
• 0.56 - 0.79 N m
• 5 - 7 in-lbs
Please do not overtighten the screws.
ower
TS-3
1
15
16
17
RLY 1
Power
WARNINGS:
The DC input terminals shall be supplied from a source providing electrical
isolation between the input and output, in order to meet the applicable safety
requirements of IEC 61010-1.
All field wiring must have insulation suitable for rated voltages.
Important!
Before applying power to the SITRANS LUT400 for the first time, ensure any
connected alarm/control equipment is disabled until satisfactory system operation
and performance is verified.
7ML19985XU81
TRANSDUCER
DISCRETE INPUTS
2
3
4
5
6
7
8
1
terminal blocks
RELAYS
18
19
20
21
RLY 2
RLY 3
display cable
SITRANS LUT400 (HART)
INPUT
POWER
9
10
11
12
13
2
DC
SYNC
L1
L2/N
0.25 A, 250 V
100 - 230 V
SLOW BLOW
mA OUT/HART
BATTERY
BR2032
DISPLAY
22
23
24
PAS
ACT
SERVICE
mA HART
connection
power
supply
14
battery
USB
connection
Page EN-11