Putting into Operation
Press the brake release lever (fig. 18) forwards and downwards fully.
The control will recognize that the brake has been released and automatically
deactivate the drive function.
The error indicator shows 2 or 3 flashing LEDs and emits an acoustic signal when the
brake is released.
70
Fig. 17:
Brake release bolt
Fig. 18:
Brake release lever
Attention!
If the control shows no error signal when the brake is
released, the brake setting must be checked by the
healthcare provider.
Attention!
Once you have disengaged the brake release lever for the
pushing function, all brake systems are switched off.