Connections on the rear panel of the camera control unit
Connections
Description
RGB/SYNC 24
For feeding a high-quality analog RGB signal to
a monitor or other peripherals through a cable
with a 15-pin HD connector.
S-VIDEO OUT
Outputs for Y/C-signals for connecting a mon-
(1 + 2) 25
itor or other peripherals via cables with mini
DIN plugs
VIDEO OUT
Outputs for Composite Video signals for con-
(1 + 2) 26
necting a monitor or other peripherals via
cables with BNC plugs
REMOTE 27
Connector for remote control of peripherals
(video printer, video recorder or other docu-
mentation devices) via cables with CINCH
plugs
Equipotentialization
Connector for equipotentialization leads
connector 28
according to DIN 42801
Fuse holder 29
For standard fuse T 0.5 A L/250 V~
(Art. no. TA020394)
Mains connector 30
Connector for power cord IEC 320, see Acces-
sories/Spare parts
4.
Preparation and setup
Non-compliance with the following rules will result in complete exclusion
of liability on the part of Aesculap.
When setting up and operating the product, adhere to
– national regulations for installation and operation,
– national regulations on fire and explosion protection.
Note
For the safety of patients and users it is essential that the mains power cord
and, especially, the protective earth connection are intact. In many cases
defective or missing protective earth connections are not registered imme-
diately.
Connect the device via the potential equalization terminal at the rear
panel of the device to the potential equalization system of the room
used for medical purposes.
Ensure that all devices operated in the vicinity comply with their cor-
responding EMC requirements.
Note
The potential equalization lead can be ordered from the manufacturer as
art. no. GK535 (4 m length) or TA008205 (0.8 m length).
Note
Arrange the medical cart outside the sterile area.
Note
Measure the cable length so that there is sufficient freedom of movement
without endangering the sterility.
Note
Make sure access to the power connection is unobstructed.
4.1
First use
Risk of injury and/or product malfunction due to
incorrect operation of the electromedical system!
Adhere to the instructions for use of any medi-
WARNING
cal device.
There are three different connection variants, depending on system
requirements.
If a parallel video image from the documentation device is not
required, always connect the components as shown under variant 1.
If a parallel video image from the documentation device is required on
a monitor, connect the components as shown under variant 2 or 3.
Always connect the video output (Video out) of one device to the video
input (Video in) of the next device.
1. variant: Connecting the camera, monitor and one documentation
device to the camera control unit (no parallel video images)
Camera
Monitor
Video in
Video out
Video cable 1
Select signal or type of video cable according to the common signal
standard of the camera and monitor, always choosing the highest-
quality signal option (video signals in descending order of quality: RGB,
S-Video (Y/C), Composite Video (FBAS)).
External device
e.g. recorder
Video out
Video in
Video out
Video cable 2
5