3-Chip video camera system DAVID 3 PV430/PV432/PV435
6.5
Cleaning/disinfection
Product-specific safety instructions for the reprocessing procedure
Risk of electric shock and fire hazard!
Unplug the device before cleaning.
DANGER
Do not use flammable or explosive cleaning or
disinfecting solutions.
Ensure that no fluids will penetrate the product.
Validated cleaning and disinfection procedure
Validated procedure
Special features
Wipe disinfection for
electrical devices with-
out sterilization
Camera head and cam-
era control unit after
every use
6.6
Wipe disinfection for electrical devices without sterilization
Phase
Step
I
Wipe disinfection
RT:
Room temperature
Phase I
Remove any visible residues with a disposable disinfectant wipe.
Wipe all surfaces of the optically clean product with a fresh, disposable
disinfectant wipe.
Observe the specified application time (1 min minimum).
6.7
Inspection, maintenance and checks
Remove the moisture protection cap from the camera plug and check
that the plug pins are dry.
Allow the product to cool down to room temperature.
Inspect the product after each cleaning and disinfecting cycle to be
sure it is: clean, functional, and undamaged.
10
®
Wipe all surfaces of the product with a single-use cleaning wipe.
Move the buttons and slides repeatedly during cleaning and disin-
fecting.
Dry the product with a lint-free cloth.
When using other disinfectants: Observe the concentration informa-
tion and application times of the disinfectant manufacturer.
T
t
Conc.
[°C/°F]
[min]
[%]
≥1
RT
-
Damage to the product due to inappropriate clean-
ing/disinfecting agents!
CAUTION
Risk of damage to the product due to inappropriate
cleaning/disinfecting agents!
CAUTION
Water quality
Chemical
-
Meliseptol HBV wipes 50 % Propan-1-ol
6.8
Packaging
Follow the instructions for use for the applied packaging and storage
systems (e.g. instructions for use TA009721 for Aesculap Eccos storage
system).
Insert the product in its proper position in the Eccos holder, or put it
on a tray in such a way that the product is protected against damage.
Pack trays appropriately for the sterilization process (e.g. in Aesculap
sterile containers).
Ensure that the packaging will prevent a recontamination of the prod-
uct.
Only use cleaning/disinfecting agents approved
for surface cleaning. Follow the manufacturer's
instructions for the respective cleaning/disin-
fecting agent.
Use cleaning and disinfecting agents according
to the manufacturer's instructions which
Observe specifications regarding concentration,
temperature and exposure time.
Reference
Chapter Wipe disinfection for electrical devices
without sterilization