3-Chip video camera system DAVID 3 PV430/PV432/PV435
8.
Troubleshooting list
Malfunction
No image on the monitor
Image too dark
Image fuzzy or poorly defined
Image fogged
Noise or excessive grain in the picture
Image too bright or glary
Color bar displayed on the monitor
Unrealistic color representation on the monitor
12
®
Cause
Device not powered
Fuse blown
Wrong video input activated
Video cable defective
Video cable not connected
Camera cable not connected or defective
Light cable not connected
Light cable defective
Light source dimmed down
Endoscope optical system defective
Object out of focus
Debris on the endoscope tip
Moisture on the camera cable plug
Brightness enhancement function is active
Light metering not at its optimum setting
Camera cable not connected to the camera con-
trol unit
Automatic white balance failed
Monitor color setting misadjusted
Video cable defective
Remedy
Verify that all units are connected to mains
power and switched on
Replace camera fuse T0.5 A L/25 V~ (TA020394)
If the monitor supports several channels: Verify
that the monitor is set to the channel that is
presently used (e.g. RGB, S-Video (Y/C), Com-
posite Video (FBAS))
Replace video cable
Connect video cable
Ensure that the camera head is connected to the
camera control unit and the camera plug is dry
Verify that the light cable is connected at the
light source and the light cable input of the
endoscope
Connect a new light cable
Have repaired by manufacturer
Turn up the light source intensity
If the image is too dark even without the cam-
era: Use a different endoscope and have the
defective endoscope repaired by the manufac-
turer
Turn the focus adjustment setting ring of the
endo-lens until there is a sharp image
Clean tip with a sterile cloth and isopropyl alco-
hol (70 %)
Dry plug with gauze before plugging it in at the
camera socket
Check if the front and rear windows of the
endoscope and the front window of the endo-
lens are clean; if necessary, clean with isopropyl
alcohol (70 %)
Increase intensity of the light source
Press MODE button and select the optimum
light metering mode for the sheath diameter of
the endoscope
Connect camera cable at the camera control
unit
Start automatic white-adjustment
Reset to factory setting
Replace video cable