3-Chip video camera system DAVID 3 PV430/PV432/PV435
3.
Product description
3.1
Scope of supply
3-Chip video camera system DAVID 3 PV430/PV432/PV435 comprises:
Camera control unit
Camera head with endo-lens and camera cable
RGB connecting cable PV424
Instructions for use TA011450
3.2
Components required for operation
In addition to the camera system, the following components are required
for an endoscopic intervention:
Power cord
Endoscope
Light source
Light cable
Monitor
3.3
Intended use
The 3-chip video camera system DAVID 3 PV430/PV432/PV435
(DAVID = Digital Aesculap Videoendoscopic Image Display) is used in
medical endoscopy, together with an endoscope, a light source and a
monitor.
It serves to visualize the endoscopic image for display on a color monitor.
Indication
The product is indicated for supporting diagnosis and treatment in imag-
ing endoscopy.
Contraindication
No specific contraindication can be provided for use can be stated due to
the universal suitability of the product for endoscopy. The use of the prod-
uct is contraindicated if endoscopic procedures are contraindicated. In
general, all uses that include CF conditions are contraindicated.
4
®
3.4
Operating principle
Endoscope 10 visualizes the operation site and endoscopic lens 11
focuses this image onto the three CCD sensors, where it is converted into
electrical signals that can be further amplified. The image brightness is
automatically adjusted to the light intensity (autoshutter).
Through camera cable 17, the image signals are transferred to camera
control unit 1, where they are further amplified for display on the monitor.
The video signal processed in this way can then be picked up at the video
output sockets for connecting a monitor, video recorder, video printer or
any other digital documentation or archiving system.
For areas that are particularly difficult to illuminate, the camera system
offers a dynamic electronic brightness enhancement function (GAIN),
which automatically adapts to different light conditions, and a choice of
three mode settings for different situations during application.
The video camera system is also equipped with an automatic white bal-
ance system for optimal color rendition.
Control elements and indicators on the front panel of the camera con-
trol unit
Control elements /
Description
Displays and indi-
cators
ON/OFF switch 2
Green LED 3 above the ON switch is on when the
device is switched on.
Camera socket 9
For connecting the camera plug at the end of the
camera cable
Pushbutton white
For precise white-balancing, ensuring realistic
balance (WHITE) 4
rendition of all other colors
Pushbutton light
For setting the optimal light metering mode for
metering (MODE) 5
the sheath diameter of endoscope in use.
Color bar diagram
Displayed on the monitor when camera control
unit 1 is switched on without camera cable 17
being connected