8. Select a water temperature that aligns with your target patient temperature.
N N o o t t e e - - Allow the water to flow from the controller into the thermal transfer device until full.
9. Tap Confirm.
S S e e l l e e c c t t i i n n g g a a t t h h e e r r a a p p y y m m o o d d e e
You can select from one of three therapy modes and tap Confirm:
Automatic therapy
Manual therapy
Monitor non-therapy
For mode descriptions, tap the Help button.
W W A A R R N N I I N N G G
• Always check the integrity of the patients skin and temperature according to hospital protocol when using the A A l l t t r r i i x x
system.
• Always use Stryker accessories. Only IEC 60601-1 equipment shall be hooked to the patient temperature ports. Failure
to comply with these instructions may invalidate any or all warranties and may negatively affect the products EMC
performance. This also protects the product from cardiac defibrillation.
• When using the temperature controlled Automatic therapy mode for warming (min, med, or custom), switching to other
modes, changing the target patient temperature, or changing the therapy selection may impact the overall benefit of
therapy.
C C A A U U T T I I O O N N
• Explosion risk. This product is not suitable for use in the presence of flammable anesthetic mixture with air or with
oxygen or nitrous oxide other than nasal or mask type.
• Always make sure that there are no water leaks before starting a defibrillation.
• Always monitor the patient for shivering, temperature, signs of intolerance, and skin condition when using this product.
S S t t a a r r t t i i n n g g A A u u t t o o m m a a t t i i c c t t h h e e r r a a p p y y m m o o d d e e
In Automatic mode, the therapy cools or warms the patient to a selected patient target temperature. The product in
automatic mode continually measures the patient temperature and automatically adjusts the water temperature until the
selected patient target temperature is achieved. After the selected patient target temperature is achieved, the product
maintains this temperature for the duration of the therapy.
To start Automatic therapy mode:
1. Prepare the thermal transfer devices for therapy.
2. See Filling a thermal transfer device (page 26).
3. Apply the thermal transfer device to the patient.
4. Connect the reusable adapter cable to port A or port B on the product. Make sure that the probe is fully seated.
5. Apply the sensing end of a patient probe to the patient based on your hospital protocol and secure the product to reduce
the risk of accidental dislodgment.
6. Connect the patient temperature probe to the reusable adapter cable. See Selecting and connecting a temperature
probe (page 20).
7. Tap to Confirm the current patient temperature.
8. Tap the Automatic therapy mode button.
9. Select the target patient temperature.
10. See Setting or editing the cooling rates (page 28) or Setting or editing the warming rates (page 28)
8001-109-005 Rev AA.3
27
EN