Herunterladen Inhalt Inhalt Diese Seite drucken

Stryker Altrix 8001 Bedienungsanleitung Seite 20

Präzisions-temperaturmanagementsystem
Vorschau ausblenden Andere Handbücher für Altrix 8001:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 197
I I c c o o n n , , y y e e l l l l o o w w
N N a a m m e e
Normothermia
deviation
Therapy pause
Battery low
Patient
temperature
output (A or B)
Remove from use
(RFU)
Power loss
N N o o t t e e
• If any of the alarm conditions persist, call maintenance.
• If page indicators appear on the alarm screen, there are multiple active alarms. The highest alarm is displayed. Tap Next
or Back to view the active alarms.
C C o o n n t t a a c c t t i i n n f f o o r r m m a a t t i i o o n n
Contact Stryker Customer Service or Technical Support at: 1-800-327-0770.
Stryker Medical
3800 E. Centre Avenue
Portage, MI 49002
USA
N N o o t t e e - - The user and/or the patient should report any serious product-related incident to both the manufacturer and the
Competent authority of the European Member State where the user and/or patient is established.
To view your operations or maintenance manual online, see https://techweb.stryker.com/.
Have the serial number (A) of your Stryker product available when calling Stryker Customer Service or Technical Support.
Include the serial number in all written communication.
EN
16
A A l l a a r r m m p p r r i i o o r r i i t t y y
M M e e s s s s a a g g e e
a a n n d d d d e e l l a a y y
Low
Patient
temperature is
outside of 36.0° C
(96.8° F) to 38.0°
C (100.4° F)
Medium
Therapy is
currently paused
Low
Battery is low
Low
Patient
temperature
output is
inaccurate on the
external device,
or outside
supported range
Medium
The system has
powered off due
to a malfunction
Medium
Not applicable
T T h h e e r r a a p p y y
C C h h e e c c k k
i i n n t t e e r r r r u u p p t t e e d d
No
Check patient
condition, placement
of thermal transfer
devices, and all
connections
Yes
To resume, press
and hold the play /
pause for 2 seconds
No
Maintenance is
recommended. If
battery is not
replaced, the product
may not function on
the next startup.
No
Check output
adapter cable
connection. Tap
Confirm to reactivate
the output port.
Yes
Remove the product
from use
immediately. Notify
the appropriate
personnel.
Yes
Check power cord
connection
8001-109-005 Rev AA.3

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis