7. Tap Confirm.
8. Select the water target temperature of 25.0° C (77.0° F).
9. Tap Confirm.
10. Allow the controller to run for 2 minutes to purge air from the system.
11. Allow the controller to run for 5 minutes.
N N o o t t e e - - The timer will run on the main display, follow the current therapy duration timer.
12. After seven minutes, turn the controller off by pressing and holding the Stand-by button for two seconds.
13. Unplug the power cord from the wall outlet.
14. Place the controller over a floor drain or drain pan.
15. Remove the reservoir. Pull forward at an angle, and lift out the reservoir.
16. Remove the bottom hose end from the hydraulic connector adapter in the reservoir lid by pushing down on the collar.
17. Empty water from the reservoir, dispose of the water per hospital protocol.
18. Pull up on the controller drain plug to open the drain.
19. Make sure that all three hoses remain raised above the connection ports for draining.
20. Allow the controller and hoses to drain for a minimum of two minutes.
21. Push down on the controller drain plug to close the drain.
22. Empty the water from the drain pan, if applicable.
23. Wipe the inside and outside of the reservoir and reservoir lid, with a dry, soft, lint free cloth.
24. Place the reservoir into the controller.
25. Disconnect and store the service tool adapter hoses from all three of the hoses. (If applicable, when used with colder
style hoses.)
26. Store the power cord, cables, and hoses.
EN
48
1
8001-109-005 Rev AA.3