6. Commissioning, operation

6.5.4 Content of the SETUP menu

EN
MEASURING operating mode
1
To measure working or process pressures
For comparative measurements and/or calibrations (without data recording) of mechanical
1)
and electrical pressure measuring instruments (supply and display of the test item through
the CPH7650)
For further information, see chapter 6.6.1 "MEASURING mode" and 6.6.2 "MEASURING
mode (with test item)"
CALIBRATION operating mode
2
For on-site calibration of mechanical
PC). In this case the data sets (for up to 16 test items, each with up to 32 test steps including
date and time) are recorded within the CPH7650.
For more information please refer to chapter 6.6.3 "CALIBRATION mode"
SWITCH TEST operating mode
3
For the easy checking of pressure switches, including automatic calculation of the switch
hysteresis
For more information please refer to chapter 6.6.7 "SWITCH TEST mode"
Functions
4
Tare: Offset correction of the reference pressure value
Min/Max: Minimum/Maximum memory
Alarm: Min/Max alarm (visual and audible)
Filter: Damping/smoothing of the reference sensor signal
For more information please refer to chapter 6.7.1 "SETUP additional menu items: Functions"
1)
For mechanical dial instruments, the test item's measured value must be entered via the numeric keypad.
38
.Configuration
1
MEASURING
2
CALIBRATION
3
SWITCH-TEST
4
Functions
5
CPH-Info
6
ReferenceSensor
7
Ref. Sensor-list
8
CPH-Configuration
9
Interface
10
CLEAR CalProg
11
CPH switch off
1)
and electrical pressure measuring instruments (without
WIKA operating instructions, model CPH7650