6. Commissioning, operation
6.7

Setup menu

The instrument features 3 operating modes. MEASURING / CALIBRATION / SWITCH TEST, which offer the maximum
convenience in relation to its application.
To power the test items and to read their measuring signals, there are electrical inputs and outputs which can be protected
from adverse conditions in the field by optional captive protective caps.
EN
Operating modes
MEASURING and
CALIBRATION
SWITCH TEST
With respect to the transfer of data to a PC, the pressure calibrator has a RS-232 and USB interface, selectable via menu. For
further information, see chapter 6.9.6 "Interface".
1. With [SETUP], access the main menu "Configuration".
2. With [▲] or [▼], select the desired menu item.
3. Confirm with [SELECT].
Menu item is displayed.
MEASURING operating mode
1
To measure working or process pressures
For comparative measurements and/or calibrations (without data
recording) of mechanical
instruments (supply and display of the test item through the CPH6000)
→ See chapter 6.8.1 "MEASURING mode"
CALIBRATION operating mode
2
For on-site calibration of mechanical
instruments (without PC). In this case the data sets (for up to 16 test items,
each with up to 32 test steps including date and time) are recorded within
the CPH6000.
→ See chapter 6.8.2 "CALIBRATION mode"
SWITCH TEST operating mode
3
For the easy checking of pressure switches, incl. automatic calculation of
the switch hysteresis
→ See chapter 6.8.3 "SWITCH TEST mode"
Functions
4
Tare: Offset correction of the reference pressure value
Min./Max.: Minimum/Maximum memory
Alarm: Min./Max. alarm (visual and audible)
Filter: Damping/smoothing of the reference sensor signal
→ See chapter 6.9.1 "Functions"
CPH info
5
General CPH6000 instrument data
Calibration data for the electrical measuring inputs
Firmware number
Serial number of the instrument
→ See chapter 6.9.2 "CPH info"
1) For mechanical dial instruments, the test item's measured value must be entered via the numeric keypad.
28
Function
In the MEASURING (with test item) and CALIBRATION operating modes, the measured values of both
the reference pressure sensor and the test item, as well as their deviation, are displayed in both current
pressure units and in %.
In this way the operator is immediately informed online whether the test item meets the class accuracy
or not.
The calibration data in CALIBRATION mode are stored internally and can later be transferred onto
printable certificates through software (WIKA-Cal).
In the operating mode SWITCH TEST, pressure switches can be checked and an automatic calculation
of the switch hysteresis can be performed.
1)
and electrical pressure measuring
1)
and electrical pressure measuring
WIKA operating instructions, pressure calibrator model CPH6000
.Configuration
1
MEASURING
2
CALIBRATION
3
SWITCH-TEST
4
Functions
5
CPH-Info
6
ReferenceSensor
7
Ref. Sensor-list
8
CPH-Configuration
9
Interface
10
CLEAR CalProg
11
CPH switch off