6. Commissioning, operation
Battery level:
8
Current battery level
At 10 % the low battery charge warning, "low BAT" is shown in the display.

6.9.6 Interface

The interfaces can be switched on and off. In order to extend the battery life, the interfaces should be deactivated when not in
use.
With the [▲] or [▼] button, select the interface.
Confirm selection with [SELECT].
With the [◀] or [▶] buttons, change between the interface and baud rate.
RS-232 interface
For the RS-232 interface there is a choice of 3 different baud rates with fixed 8 used data bits,
no parity and 1 stop bit.
USB interface
The USB interface can only be switched on and off.

6.9.7 CLEAR CalProg

If the SETUP menu item CLEAR CalProg is selected with the cursor and then the [SELECT] button is pressed 2 x, then all
stored calibration data will be deleted or reset.
The status of the deletion process is shown at the left of the lower info screen.

6.10 Working with WIKA-Cal

WIKA-Cal calibration software serves for the creation of calibration certificates or logger protocols for pressure measuring
instruments.
The WIKA-Cal calibration software enables both reading the logger data stored in the hand-held as well as online calibrations
together with a PC. The scope of software functions depends on the selected licence.
Further information on this, and also the operation, can be found in the instruction manual of WIKA-Cal.
WIKA operating instructions, pressure calibrator model CPH6000
..Interface
USB
ON
RS232
9600
(8N1)
41