6. Commissioning, operation
Current date
5
The date of the integrated real-time clock in the "SETUP\CPH configuration" menu, which is later marked on the
calibration certificate. This value must be correct for the following measurement process in calibration mode and/or be
adjusted accordingly in the "SETUP/CPH configuration" menu.
→ See chapter 6.9.5 "CPH configuration".
Calibration date
6
Calibration date for the electrical measuring inputs of the CPH6000 (year/month/day)
Following the status messages, the display returns to the screen for the last-selected operating mode.
Data loss!
If the charge level shows insufficient battery capacity, charge the pressure calibrator before starting work.
When the rechargeable battery is almost completely discharged, the message "low BAT" appears in the
display. To avoid a data loss, the instrument must be charged immediately. With 0 % battery level, the equipment
automatically switches itself off and must then be recharged using the power supply unit.
All data in the internal memory will be lost when the battery has 0% battery capacity!

6.1.2 Switching the pressure calibrator off

The instrument is switched off using the "CPH switch of" menu item in
the first submenu.
1. For this, press [SETUP] once.
2. With [▲] or [▼] select the menu item Switch off CPH.
3. Confirm with [SELECT].
The CPH6000 switches off.
WIKA operating instructions, pressure calibrator model CPH6000
.Configuration
MEASURING
CALIBRATION
SWITCH-TEST
Functions
CPH-Info
ReferenceSensor
Ref. Sensor-list
CPH-Configuration
Interface
CLEAR CalProg
Switching off
CPH switch off
EN
21