11. Specifications
Output signal
Signal type
Current
EN
Voltage
Connection method
Operating conditions model CPT6000
Medium temperature range
Ambient temperature range
Storage temperature range
Relative humidity
Ingress protection
10.4 Power supply unit
Power supply unit
Permissible ambient conditions
Operating temperature range
Storage temperature range
Relative humidity
10.5 Test cables
Test cable set (red and black)
Scope
Operating temperature
Dimensions
Nominal voltage
Nominal current
50
4 ... 20 mA
■
0 ... 20 mA
■
DC 0 ... 1 V
■
DC 0 ... 2 V
■
DC 0 ... 5 V
■
DC 0 ... 10 V
■
2-wire
■
3-wire
■
-20 ... +80 °C [-4 ... +176 °F]
-20 ... +80 °C [-4 ... +176 °F]
-40 ... +85 °C [-40 ... +185 °F]
0 ... 95 % r. h. (non-condensing)
IP54
■
IP65 (with cable connected)
■
0 ... 45 °C [32 ... +113 °F]
-40 ... +85 °C [-40 ... +185 °F]
5 ... 95 % r. h. (non-condensing)
2 pairs of measuring lines
■
2 pairs of clamp connectors
■
2 pairs of crocodile clips
■
-15 ... +70 °C [-30 ... +158 °F]
4 x 500 mm [0.16 x 19.69 in]
4 x 300 mm [0.16 x 11.8 in]
DC 60 V
■
AC 30 V
■
Measuring lines
Crocodile clips
WIKA operating instructions, pressure calibrator model CPH6000
32 A
5 A