6. Commissioning, operation
Defining the first test point
Calibration
1
Number of the calibration and therefore the test item (up to 16 calibrations,
each with up to 32 test steps, can be predefined and then recorded)
Test item
2
Test item type and test item measuring signal
0 ... 20 mA / 4 ... 20 mA / 0 ... 1 V / 0 ... 2 V / 0 ... 5 V / 0 ... 10 V / or
mechanically for pointer instrument
Test point
11
Number of the test point x
Set
13
Value of the test point x (input via numeric keypad)
(Test point x+1 and x-1 accessible via [▲] or [▼])
The CPH6000 is only suitable for measuring DC voltage and DC current signals.
First test point defined
Example here:
Test point no. 1 = 0 bar
Pressure unit of the test point
6
Unit and resolution (submenu)
→ See chapter 6.7.1 "Display: Units and resolution"
Test point
11
No. 1
Set
13
Test point no. 1 = 0 bar
Defining the second test point
With [▲] or [▼], select the test point.
Call up [▶] 2nd test point.
Input of the values via numeric keypad.
Confirm by pressing [ENTER].
With [◀] back to the previous test point.
Defining further test points
Further test points are performed as described in Defining the second test
point.
It is possible to prepare up to 16 calibrations, each with up to 32
test steps.
ID no.
3
IDENT number of the test item
Tag no.
4
Measuring location number of the test item
MRS / MRE
5
Start of measuring range and end of measuring range of the test item currently to be calibrated
WIKA operating instructions, pressure calibrator model CPH6000
..Calibration
Calibration:
2
Testitem:
4-20 mA
ID no:
123456
456789
Tag no:
R.-Start:
0.000
Range-End:
6.000
Unit:
Class:
0.50
%FS
Pressuretype:
Medium:
Gas
PowerSupply:
24V
Testpoint:
Hold time:
Set:
True:
0.00000
Calibration program
..Calibration
Calibration:
Testitem:
4-20 mA
ID no:
123456
Tag no:
456789
R.-Start:
0.000
Range-End:
6.000
Unit:
Class:
0.50
%FS
Pressuretype:
Medium:
Gas
PowerSupply:
24V
Testpoint:
Hold time:
Set:
0.00000
True:
0.00000
Calibration program
..Calibration
Calibration:
1
2
Testitem:
4-20 mA
3
ID no:
123456
456789
4
Tag no:
R.-Start:
0.000
5
Range-End:
6.000
6
Unit:
Class:
0.50
7
%FS
8
Pressuretype:
9
Medium:
Gas
10
PowerSupply:
24V
11
Testpoint:
12
Hold time:
13
Set:
14
True:
0.00000
Calibration program
1
1
EN
bar
rel
11
1
0
13
1
6
bar
rel
11
1
0
13
1
bar
rel
2
0
33