6. Commissioning, operation

6.8.3 SWITCH TEST mode

In SWITCH TEST mode, along with the reference pressure sensor data, the status and switch points of the pressure switch are
also displayed.
Accessing SWITCH TEST mode:
1. With [SETUP], access the main menu.
2. With [▲] or [▼], select menu item SWITCH TEST.
3. Confirm with [SELECT].
The submenu SWITCH TEST is displayed.
Preparing the SWITCH TEST:
The switch test is not suitable for electronic switches (e.g. PNP- or NPN
switches), but is only for mechanical, potential-free switches.
With [▲] or [▼], select menu item.
Use [◀] or [▶] to make the configuration.
When finished, confirm all settings with [SELECT].
SWITCH TEST mode is initiated.
Voltage supply for test item (on/off)
1
If no external supply is required for the test item, "OFF" should be selected
to conserve energy (also see chapter 4.2 "Electrical connections").
Unit
2
Unit and resolution (submenu)
→ See chapter 6.7.1 "Display: Units and resolution"
The calculated measured values of the two switch points and the
hysteresis can be reset by pressing the 0 button.
WIKA operating instructions, pressure calibrator model CPH6000
.Configuration
MEASURING
CALIBRATION
SWITCH-TEST
Functions
CPH-Info
ReferenceSensor
Ref. Sensor-list
CPH-Configuration
Interface
CLEAR CalProg
CPH switch off
..Switch-Test
1
PowerSupply:
OFF
2
Unit:
bar
Test item 24V/50mA
EN
37