10. Dismounting, return and disposal / 11. Specifications
To avoid damage:
1. Wrap the instrument in an anti-static plastic film.
2. Place the instrument, along with the shock-absorbent material, in the packaging. Place shock-absorbent material evenly on
all sides of the transport packaging.
3. If possible, place a bag, containing a desiccant, inside the packaging.
4. Label the shipment as carriage of a highly sensitive measuring instrument.
Information on returns can be found under the heading "Service" on our local website.
9.3

Disposal

Incorrect disposal can put the environment at risk.
Dispose of instrument components and packaging materials in an environmentally compatible way and in accordance with the
country-specific waste disposal regulations.
Do not dispose of with household waste. Ensure a proper disposal in accordance with national regulations.

10. Specifications

10.1 Basic information - Complete measuring chain

Complete measuring chain
Sensor technology
Version
Accuracy of the measuring chain
Gauge pressure
Absolute pressure
Pressure types
Gauge pressure
Absolute pressure
Vacuum and +/- measuring range
1) It is defined by the total measurement uncertainty, which is expressed with the coverage factor (k = 2) and includes the following factors: the intrinsic performance of the instrument, the
measurement uncertainty of the reference instrument, long-term stability, influence of ambient conditions, drift and temperature effects over the compensated range during a periodic
zero point adjustment.
2) Calibrated at 23 °C [74 °F] and in a vertical mounting position, pressure connection facing downwards.
WIKA operating instructions, pressure calibrator model CPH6000
1 reference pressure sensor CPT6000 (interchangeable without tools)
External operation via 1.1 m [3.6 ft] sensor connection cable (plug-and-play)
Up to ten data sets or reference pressure sensors can be programmed per CPH6000.
1) 2)
0.025 % FS
0.1 % FS
0.025 % FS
0 ... 6,000 bar [0 ... 87,000 psi]
0 ... 0.25 bar abs. to 0 ... 25 bar abs.
[0 ... 4 psi abs. to 0 ... 360 psi abs.]
-600 ... +600 mbar to -1 ... 39 bar
[-10 ... +10 psi to -14.5 ... 550 psi]
≤ 1,000 bar [≤ 14,500 psi]
> 1,000 bar ... ≤ 6,000 bar [> 14,500 psi ... ≤ 87,000 psi]
≤ 25 bar abs. [≤ 360 psi abs.]
EN
45