6 Commissioning, operation
6.2.1

Main menu functionality

There are three options on the main menu, "CONFIG", {active display} and "MORE". The main
menu is home for the menu display.
GB
6.2.1.1 Setting the active display
The active display is indicated by the center option on the main menu, pressing the F2 key will
toggle the active display.
6.2.1.2 Setting active display parameters
To set the parameters of the active display use the "CONFIG" option to get to the display
configuration menu.
Here the "SELECT" option will toggle through the choices for each parameter. The first parameter
is "MODE". Since current and switch test modes all use the same jacks, two of these functions
cannot be used concurrently. The ability to select certain functions is limited based on what is
already selected in another active display. The "NEXT" option is used to change to the second
parameter. Only RTD (resistance thermometer) and pressure modes have a second parameter,
RTDs can be read in Celsius or Fahrenheit and pressures can be read in 17 engineering units.
Within any single display sector, the following modes are available:
P[1] = pressure on left side sensor
P[2] = pressure on right side sensor
P[1] ST = switch test with left side sensor
P[2] ST = switch test with right side sensor
mA = current measurement
RTD = resistance thermometer measurement (in the event a resistance thermometer is connected)
The following table shows which functions are available concurrently.
An "--" in a column indicates that the mode in the active display will not be available for selection if
the mode in that row is in use in any other active display.
18
P[1]
P[1]
P[2]
P[1] ST
P[2] ST
mA
RTD
P2 is only available on the model CPH65I0-S2.