Herunterladen Inhalt Inhalt Diese Seite drucken

Device Preparation - Nouvag MD11 Bedienungsanleitung

Vorschau ausblenden Andere Handbücher für MD11:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 1
6.3

Device preparation

1.
Sterilize the motor (the motor is not sterile on delivery). If the motor has already been sterilized: when
removing the motor from the sterile packaging, ensure that the sterile packaging is not damaged and
that the sterility indicator confirms sterility (if no sterility indicator is provided, the sterile packaging
must at least show the date on which the shelf life of the sterile item is due to expire).
2.
Insert the stand for the irrigation fluid into the stand holder.
3.
Plug the motor plug of the electronic motor into the motor socket.
4.
Plug the pedal plug into the pedal socket at the rear of the control unit.
5.
Attach the sterilized handpiece to the electronic motor. Press the handpiece firmly onto the electronic
motor until it clicks into position and make sure it is secure by moving it slightly in the opposite direction.
6.
Assemble the tubing set: Attach tubing set REF 1706, for the cooling of the contra angle, like described
in the following article.
Use only Nouvag tube set REF 1706, otherwise the function cannot be guaranteed.
Check the expiry date of the tubing set and ensure that the packaging is not damaged. Using
non-sterile tubing sets can result in serious infection and, in extreme cases, can be fatal.
When inserting the tubing set, watch the arrow on the cover of the pump compartment.
It indicates the flow direction of the cooling liquid.
Do not regulate the amount of irrigation fluid using the roller clamp on the tubing set; with the MD 11, this
is regulated instead using the integrated pump. For this reason, make sure to open the roller clamp as far
as it will go (please refer to 7.4.5 Setting the pump supply quantity).
A
A) Press release key for tubing set bracket (on top of the control unit) to open the pump.
B) The compartment with the integrated tubing bracket opens.
C) Place the tubing set into the tubing bracket provided in such a way that the part of the tubing set
with the spike exits the pump towards the rear of the device. Make sure the tubing is secure.
D) With the tubing set inserted, press the compartment downwards until it clicks into place.
8
B
MD 11, V2.0, Operating Instructions, REF 31665, V22/17
C
MD 11
D

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis