Device preparation

1.
Sterilize the motor (the motor is not sterile on delivery). If the motor has already been sterilized: when
removing the motor from the sterile packaging, ensure that the sterile packaging is not damaged and
that the sterility indicator confirms sterility (if no sterility indicator is provided, the sterile packaging
must at least show the date on which the shelf life of the sterile item is due to expire).
2.
Insert the stand for the irrigation fluid into the stand holder.
3.
Plug the motor plug of the electronic motor into one of the motor sockets.
4.
Where appropriate, plug the motor plug of a second electronic motor into a motor socket.
If you want to use a handpiece with the 80,000 rpm high speed motor and change ist parameter,
than the motor has to be plugged in one of the motor sockets. Otherwise there will be only
displayed the values of the standard 50,000 rpm motor.
The same is with the OTO-Drill handpiece with its integrated 16,000 rpm motor.
Note: Always plug in the motor first, than change the parameters of the selected handpiece.
5.
Plug the pedal plug into the pedal socket at the rear of the control unit.
6.
Attach the sterilized handpiece to the electronic motor. Press the handpiece firmly onto the electronic
motor until it clicks into position and make sure it is secure by moving it slightly in the opposite
direction.
7.
Assemble the tubing set: Decide whether you will use tubing set Ref. 6024, for the cooling of a single
contra angle or if you need to use tubing set Ref. 6025 with an integrated 3 way cock for cooling two
contra angles at the same time when two motors are used.
Use only Nouvag tube sets Ref. 6024 and 6025, otherwise the correct function cannot be guaranteed.
Check the expiry date of the tubing set and ensure that the packaging is not damaged.
Using non-sterile tubing sets can result in serious infection and, in extreme cases, can be
fatal.
When inserting the tubing set, watch the arrow on the cover of the pump compartment. It indicates
the flow direction of the cooling liquid.
Do not regulate the amount of irrigation fluid using the roller clamp on the tubing set; with the
HighSurg 30, this is regulated instead using the integrated pump. For this reason, make sure to open
the roller clamp as far as it will go (please refer to 7.4.5 Setting the pump supply quantity).