Herunterladen Inhalt Inhalt Diese Seite drucken

Device Preparation - Nouvag Vacuson 60 LP Bedienungsanleitung

Liposuction system
Vorschau ausblenden Andere Handbücher für Vacuson 60 LP:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Vacuson 60 LP
6.6

Device preparation

1. Prior to use, all sterilizable parts (e.g. Silicone
tubes, Cannulas, Conform cannula handpiece, han-
dlebar, char lid and bottles) must be sterilized.
2. Attach secretion jars with mounted and locked lid
to the device.
3. Lay the VARIO foot switch with pedal on the floor
and connect the power plug with the foot switch
socket.
4. Open the hood of the tube compartment and in-
sert the tubing set. Make sure the flow direction
corresponds with the arrows on the tube com-
partment.
Operating Instructions, Vacuson 60 LP, REF 31793, V34/17
Adjust the clamps at the left and at the right of the
infiltration pump to fix the tube set in place. There-
fore close the tube compartment and gently adjust
the clamps by twisting the knurled wheels into the
appropriate position.
5. Hang bottle with infiltration solution on the stand
and insert the spike through the diaphragm at the
bottle top. Open the bleed valve.
Attach infiltration cannula to the to the cannula
adapter. Open the roller clamp all the way. The
amount of infiltration liquid is adjusted by the
pump (INFILTRATION).
EN
13

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis