Herunterladen Inhalt Inhalt Diese Seite drucken

Tubing Set, Ref 1706; Handpiece Cradle - Nouvag MD11 Bedienungsanleitung

Vorschau ausblenden Andere Handbücher für MD11:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 1
8.3

Tubing set, REF 1706

Single-use tubing set 1706 may not be reused.
Tubing sets must be disposed of properly after use!
Do not use tubing sets when packaging is already open or damaged!
Do not use tubing sets when expiry date has passed.
Use only Nouvag tubing sets with REF 1706!
Sterility cannot be guaranteed by reusing and re-sterilization of tubing sets. The characteristics of the
material change in a manner that can result in failure of the system. This may result in serious infections
or even patient death worst-case.
8.4

Handpiece cradle

Soiled handpiece cradles are cleaned using neutral or alkaline cleaning agents in the recommended
concentration. It can be sterilized in accordance with the same instructions as for electronic motor 21.
26
MD 11, V2.0, Operating Instructions, REF 31665, V22/17
MD 11