7.8

Configuration Menu

In the Configuration Menu the user can customize the device after his favor. The parameters are organized
in two levels. The following information and parameters are stored in there.
• Software version
• Mainboard serial number
• Language selection, DE/EN
• Display illumination
• Number of programs
• Power on, at last program
• Operating hours MD 11
• Operating hours Motor
• Operating hours of irrigation pump
Be cautious when changing parameters. Unusual behavior of instruments while operating may provoke
false reactions and jeopardize the patient.
Every setting and the new behavior of the instrument has to be verified and tested.
1. Access to Configuration Menu:
• Press "HP" and "Speed ▼" key for about 3 sec. until you hear a tone.
On the display the first position of the configuration menu shows up.
The arrows at the beginning of the lines indicate
the operation in the Configuration Menu.
2. Selection of parameters:
• Press "HP" or "PR" to choose the desired parameter.
• Press "Speed ▼" to select the parameter.
• Change the value by pressing "HP" for up or "PR" for down.
• To confirm the changes made, press "Speed ▼" for about one sec., until you hear a short tone.
• To abort the settings made, press "Speed ▼" briefly. The setting returns to its previous value.
3. Exit from Configuration Menu:
• To leave the Configuration Menu press "HP" and "PR" for 3 sec. until you hear a tone.
Parameter Level 1
Group/Parameter
Software/Version
Hardware/Serial number MB
Language, 0 = DE, 1 = EN
Backlight/brightness (0. .10)
Number of programs
Power on at last program
Operating hours/Control Unit
Operating hours/Motor
Operating hours/Pump
Error memory/ 1 – 8
MD 11, V2.0, Operating Instructions, REF 31665, V22/17
Rights
Factory
read
VX.XXXX
read
XXXXXXXXXXX Shows serial number of the main board
read/write
1
read/write
9
read/write
4
read/write
Yes
read
0
read
0
read
0
read
0
• Error log
• Activation of available handpiece
• Speed limitation for each handpiece
• Range of action of AS mode
• Pump behavior
• Motor behavior
• Behavior of torque at reverse mode
• Reset to factory default
Definition
Shows current software version
Select language between German and English
Display brightness, changeable: 0, ... , 10
Select between 2 and 6 available programs
Display of the last program after switching device on.
Shows operating hours of Control Unit
Shows operating hours of motor 1
Shows operating hours of pump
8 Error messages in chronological order.