6.5

Device preparation

1. Prior to use, all sterilizable parts (e.g. Silicone
tubes, Cannulas, Conform cannula handpiece, han-
dlebar, char lid and bottles) must be sterilized.
2. Attach secretion jars with mounted and locked lid
to the device.
3. Lay the Vario foot switch with pedal on the floor
and connect the power plug with the foot switch
socket.
4. Move the lever counter clockwise to open the tube
compartment and insert the tubing set. Make sure
the flow direction corresponds with the arrows on
the tube compartment.
8
Adjust the clamps at the left and at the right of the
infiltration pump to fix the tube set in place. The-
refore close the lever of the pump and adjust the
clamps by pressing them with the thumb in and
down at the same time. Press gently against the
tube to make sure the flow of liquid is not com-
promized.
5. Hang bottle with infiltration solution on the stand
and insert the spike through the diaphragm at the
bottle top. Open the bleed valve.
Attach infiltration needle to the Luer lock connec-
tion of the tube set. Open the roller clamp all the
way. The amount of infiltration liquid is adjusted
by the pump (INFILTRATION).
Operating instructions Vacuson 60 LP, Art. No. 31793, V46/13