MD 11
2

Explanation of symbols

Important information
Do not use if the packaging is damaged
Warning
Manufacturer
The device is designed for intermittent
duty operation at "1 min ON/3 min OFF"
at 4 cycles, after 15 min. brake.
Type B applied part
Applied parts is the instrument
Do not reuse
Biohazard
Motor
Pedal
Protection against contact and continual
submerging.
Date of manufacturing
3

Safety information

Your safety, the safety of your team, and of course that of your patients is very important to us. It is
therefore essential to bear the following information in mind:
Every use of the MD 11 different to the product description defined in chapter "Intended use and operation",
causes risks for patients and trained personnel. If physical examinations and therapies are carried out
without use of the devices then the devices must be removed from the place of treatment. Avoid any
connection or close adjacency to other devices.
3.1

EMC Manufacturer's Declaration of Conformity

The use of (RF) Radio Frequency emitting devices and equipment as well as the occurrence of negative en-
vironmental factors in the close area of the MD 11 may cause unexpected or adverse operation. Use only
accessories and cables as specified in the product description. Further observe the EMC manufacturers
declaration of conformity.
3.2

Integrated peristaltic tube pump

The integrated peristaltic pump is used to cool tissue in order to prevent damage to the tissue. It may only
be operated with watery solutions, such as 0.9 % Sodium Chloride irrigation solution or "Ringer" solution.
Supplying medication using the integrated pump is expressly prohibited.
MD 11, V2.0, Operating Instructions, REF 31665, V22/17
Autoclavable at 135°C for at least 5 min.
Suitable for thermal disinfection
Sterilized using ethylene oxide
Observe the instructions for use
Electrical and electronic devices that have
reached the end of their service life comprise
hazardous waste and may not be disposed of
together with household waste. Valid local
disposal regulations apply.
Symbol indicating the serial number with the
date of manufacture (year/month).
Symbol indicating the order number.
Symbol indicating the lot number.
Date of expiry
Protective ground
Certified by the Canadian Standards
Association (CSA)
CE symbol with notified body
EN
3