Herunterladen Inhalt Inhalt Diese Seite drucken

Configuration With Zero-Wire Teds, Version For D-Sub-Hd15; Strain Gage Full Or Half Bridge And Piezo-Resistive Sensors In 4-Wire - HBM TEDS Montageanleitung

Vorschau ausblenden Andere Handbücher für TEDS:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Fig. 2.8
View of the fully soldered plug
Important
For cable lengths above 50 m, you must loop resistors with half the bridge resistance of
the transducer into the sense leads; that is, solder one resistor each with R
the sense lead and the solder tab on the TEDS module.
4.3

Strain gage full or half bridge and piezo-resistive sensors in 4-wire

configuration with Zero-Wire TEDS, version for D-sub-HD15

The assembly procedure differs only at the end from that shown in section 4.2 starting
on page 8. So initially follow the procedure described there. But since there are no sense
leads in a 4-wire circuit, two each of the four connections on the top of the TEDS module
must be jumpered (Fig. 2.9 and Fig. 2.10).
TEDS
INSTALLATION AND CONNECTION
/2 between
B
11

Werbung

Inhaltsverzeichnis
loading