The range to be specified for the TEDS always goes from 0% to 100%, meaning that if the
lowest value is -200 N, for example, it corresponds to 0% and not -100%. Even if the
highest value in this case were 0 N, it would correspond to 100%.
Parameter
Domain Parameter
of the Calibration
Curve
Calibration Curve
Segments
1. Starting Domain
Value of Segment
Calibration Curve
Polynomial
1. Power of
1)
Domain Value
1. Polynomial
Coefficient
2. Power of
Domain Value
2. Polynomial
Coefficient
3. Power of
Domain Value
3. Polynomial
Coefficient
TEDS
TEDS TEMPLATES
Value
Unit
Physical
1
Elements
0
%
3
Elements
1
1.01946E
-02
2
9.35835E
-09
3
-4.95693
E-11
Requires
Explanation
rights at
level
CAL
Possible values:
Electrical or Physical.
CAL
Number of segments
specified.
CAL
First value in percent of the
(full) measuring range to
which the polynomial is
applied.
CAL
Number of coefficients to be
specified for the range.
CAL
Power for the first polynomial
element (in the example x
CAL
Coefficient for the first
polynomial element (in the
example x
CAL
Power for the second
polynomial element (in the
example x
CAL
Coefficient for the second
polynomial element (in the
example x
CAL
Power for the third polynomial
element (in the example x
CAL
Coefficient for the third
polynomial element (in the
example x
1
).
).
1
2
).
).
2
3
).
).
3