HBM template: Inductive Displacement Transducer
Parameter
Transducer Electrical
Signal Type
Minimum Distance
Maximum Distance
Minimum Electrical
Value
Maximum Electrical
Value
Mapping Method
Bridge Type
Transducer Response
Time
Excitation Level
(Nominal)
Excitation Level
(Maximum)
TEDS
TEDS TEMPLATES
1)
Value
Unit
Bridge
Sensor
0.000
mm
E+000
5.000
mm
E+001
0.000
V/V
80.000m
V/V
Linear
Full
1.00000
s
00u
2.5
V
10.0
V
Requires
Explanation
rights at
level
ID
This entry cannot be
changed.
CAL
The difference between
these values is the nominal
displacement according to
CAL
the data sheet.
CAL
The difference between
these values is the
sensitivity according to the
data sheet or test record, or
from the calibration.
CAL
The difference between
these values is the
sensitivity according to the
data sheet or test record, or
from the calibration.
This entry cannot be
changed.
The bridge type. The follow­
ing values are available for
selection: "Quarter" for a
quarter bridge, "Half" for a
half bridge, "Full" for a full
bridge. Some HBM trans­
ducers can be connected as
half bridges or full bridges
according to choice.
ID
Irrelevant to HBM
transducers.
ID
Nominal (rated) excitation
voltage according to data
sheet.
ID
Upper limit for the operating
range of the excitation
voltage according to the
HBM data sheet.