TEDS TEMPLATES
The terms and expression in a template are in English, as that is how they are defined in
the IEEE 1451.4 standard.
Important
Note that in some cases the number of bits for a value specified by the standard is so
small that entered numerical values are not applied exactly.
As is common practice in the templates, the following tables list prefixes to units of mea
surement, such as m (for milli-...), k (for kilo-...) or u standing for μ (micro-...) etc., in the
Value column, not the Unit column.
6.1
Basic TEDS
Parameter
Manufacturer
Model
Version letter
TEDS
TEDS TEMPLATES
1)
Value
Unit
HBM
—
WA
—
—
Requires
Explanation
rights at
level
ID
The value is stored only as a
number, e.g. 31 for HBM. Num
bers are assigned to manufac
turers in a table drawn up by the
IEEE as a supplement to the
standard. This table must be in
cluded in the hardware or soft
ware that reads the TEDS. The
entry can only be made or
changed if no other templates
have been created.
ID
The value is stored only as a
number. Numbers are assigned
to type names in a table which
can be drawn up by the respec
tive manufacturer (HBM does
this). This table must be in
cluded in the hardware or soft
ware that reads the TEDS if you
want the type name to be
shown in plain text.
ID
Possible value: a letter (A ... Z).
19