Max Output
Power
(watts)
0.01
0.1
1
10
100
7.3.4.
Guidance and Manufacturer's Declaration – Immunity – For Non-Life-Supporting Equipment
The Laser System is intended for use in the electromagnetic environment specified below. The customer or user of the device should ensure
that it is used in such an environment.
Immunity Test
IEC 60601 Test Level
Conducted RF
3 Vrms
IEC 61000- 4-6
150 kHz to 80 MHz
Radiated RF
3 V/m
IEC 61000- 4-3
80 MHz to 2.5 GHz
7.4.

FCC

This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions:
(1) this device may not cause harmful interference, and
(2) this device must accept any interference received, including interference that may cause undesired operation.
Caution: Any changes or modifications to this device not expressly approved by the manufacturer could void the user's authority
to operate the equipment.
This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to part 15 of the FCC Rules. These
limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial
environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the
instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to
cause harmful interference in which case the user will be required to correct the interference at his own expense.
This device complies with FCC RF radiation exposure limits set forth for an uncontrolled environment. This transmitter must not be co-
located or operating in conjunction with any other antenna or transmitter. The equipment should be installed and operated with minimum
distance of 20 cm between the radiator and your body.
7.5.

ISED

This device contains license-exempt transmitter(s)/receiver(s) that comply with Innovation, Science and Economic Development Canada's
license exempt RSS(s). Operation is subject to the following two conditions:
(1) This device may not cause interference.
(2) This device must accept any interference, including interference that may cause undesired operation of the device.
This Class B digital apparatus complies with Canadian ICES-003.
Radiation exposure statement:
This equipment complies with ISED radiation exposure limits set forth for an uncontrolled environment. This equipment should be installed
and operated with greater than 20cm between the radiator & your body.
7.6.

Radio Equipment Directive (RED)

The device is SAR exempt based on the operation of the device being under the SAR exemption requirements.
7.7.

Disposal

If you plan to discontinue the use of this laser system, and intend to dispose of it or any of its parts or accessories, you must observe
the applicable regional legal provisions for its disposal. You may also contact your local distributor, authorized service center, or
Customer Care for the disposal options for this laser system.
RSR-000885-000 (6)
Separation (m)
150 kHz to 80 MHz
D=(3.5/V1)(Sqrt P)
0.1166
0.3689
1.1666
3.6893
11.666
Compliance
Level
(V1)Vrms
(E1)V/m
Separation (m)
80 to 800 MHz
D=(3.5/E1)(Sqrt P)
0.1166
0.3689
1.1666
3.6893
11.666
Electromagnetic Environment – Guidance
Portable and mobile communications equipment should be
separated from the device laser by no less than the distances
calculated/listed below:
D=(3.5/V1)(Sqrt P)
D=(3.5/E1)(Sqrt P) 80 to 800 MHz
D=(7/E1)(Sqrt P) 800 MHz to 2.5 GHz
where P is the max power in watts and D is the recommended
separation distance in meters.
Field strengths from fixed transmitters, as determined by an
electromagnetic site survey, should be less than the compliance
levels (V1 and E1).
Interference may occur in the vicinity of equipment containing a
transmitter.
Separation (m)
800 MHz to 2.5 GHz
D=(7/E1)(Sqrt P)
0.2333
0.7378
2.3333
7.3786
23.333