8.
Pass the sterile connector cable from the Impella
9.
Line up the notch on the connector cable with the notch in the blue catheter plug on the front of the
Automated Impella
Controller and plug the cable into the controller.
®
10. If you have not already done so, disconnect and discard the Y connector with the red and yellow luers
from the purge tubing.
11. Connect the yellow luer on the end of the purge tubing to the yellow luer on the clear sidearm of the
Impella
®
5.0 Catheter.
12. When the controller detects that the luer is connected, it automatically begins priming the purge lumen.
Im pella 5.0 S/N: 12345
2016-07-22 10 :09
Impella
®
5.0 Circulatory Support System
5.0 Catheter off the sterile field.
®
A IC S N: 4186 A IC V 5.1
Important Step
Snapping the purge clip on
the pressure reservoir to the
connector cable is important to
prevent the tube from kinking.
9