Discard the Y Connector
After opening the purge
cassette package, disconnect
and discard the Y connector.
The Y connector is only
used with the Impella
2.5 and
®
Impella CP
Catheters.
®
Purge Solution Bottles
If the purge solution is supplied
in bottles, open the vent on the
purge fluid spike and follow the
same procedure as if supplied in
bags.
Connect Purge Disc Within
3 Seconds
The instructions for inserting the
purge disc will appear if it is not
snapped into place within
3 seconds of inserting the purge
cassette.
Close Purge Cassette Door
Once the purge cassette is
installed, be sure to close the
purge cassette door to prevent
the purge cassette from being
dislodged accidentally.
8
CASE START
CASE START
1.
Press the START NEW CASE soft button from the startup screen or plug in a new Impella
"Case Start" can also be selected by pressing the MENU soft key.
2.
The controller displays the screen shown below.
Impella 5.0 S/N: 123456
2016-07-22 10:09
INSERT PURGE CASSETTE
1.
Open the purge cassette package. Disconnect and discard the Y connector. Secure the Yellow luer to the
sterile field.
2.
Pass the purge cassette and spike off the sterile field.
3.
Press the Next soft button to continue.
4.
Open the purge cassette door by pressing the release on the left side of the controller. Insert the purge
cassette into the Automated Impella
5.
The purge cassette snaps into a molded compartment on the front of the controller. Follow the diagram
on the inside of the purge cassette door for proper placement.
6.
Slide the purge disc into the slot to the right of the purge cassette until it snaps into place. The controller
will automatically begin priming the purge cassette.
7.
Extend the purge tubing and close the purge cassette door. There is sufficient room around the edges of
the purge cassette door so that it will not pinch the purge tubing as it exits.
8.
The Automated Impella
®
CONNECT THE IMPELLA
CATHETER
®
1.
Remove the Impella
®
5.0 Catheter from its package using sterile technique and inspect the catheter,
including its connector, for damage.
2.
Remove the white connector cable from its package using sterile technique.
3.
Inspect the cable for damage, including damage to the connector pins at the controller end.
4.
Secure the black end of the cable to the sterile field.
5.
Insert the catheter plug into the connector cable socket (grey end). The tab and the slot must be aligned
during connection.
6.
Pull back on the connection to make sure that the plug has snapped into place.
7.
Snap the purge clip (located on the pressure reservoir of the clear sidearm) to the connector cable.
AIC S/N: IC4186 AIC V5.1
®
Controller.
Controller automatically begins priming the purge cassette after it is inserted.
Catheter Plug
Connector
(grey end)
Catheter.
®