Chart 1 Validated Manual Cleaning and Disinfection Procedure
Step
Step
Description
1
Contamination
Removal
2
Drying
3
Disinfection
Application
4
Manual
Disinfection
5
Final Rinse
6
Final Drying
Step Instruction
Rinse product under Cold/
Room Temperature
running tap water
removing any visible
organic material with
assistance of a soft bristle
brush
Dry the device utilizing a
dry non-shedding wipe.
Medical quality filtered air
may be utilized if available
Apply disinfectant to the
device's surface area per
manufacturer's
recommendations
While disinfectant is on the
device surface clean all
contact surfaces, joints,
mated areas utilizing a
clean soft bristle brush
Rinse product under Room
Temperature Distilled/
Filtered Water
Dry the device utilizing a
dry non-shedding wipe.
Medical quality filtered air
may be utilized if available
7
Accessories
- Cold/Room
Temperature Tap
Water
- Soft bristle brush
(Do not utilize metal
cleaning brushes)
- Non- shedding wipe
- Medical quality
filtered compressed
air
- Disinfectant
- Spray bottle or other
manual applicator
- Clean Soft bristle
brush
- Disinfectant
- Room Temperature
Distilled/Filtered
Water
- Non- shedding wipe
- Medical quality
filtered compressed
air
Duration
Until all visible soil is
removed
Until product is
visually dry
Contact time will vary
per product usage;
minimum of one (1)
minute is
recommended
Manual cleaning time
duration is complete
when the device's
surface, joints, &
crevices have been
manually cleaned
Minimum of
30 seconds
Until product is
visually dry