Herunterladen Inhalt Inhalt Diese Seite drucken

Receiving Inspection; Setup Instructions - Zimmer Air Dermatome II Bedienungsanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 68
Check the action of moving parts to ensure smooth operation throughout the intended range of motion.
The Zimmer Air Dermatome II should be returned every 12 months and the hose every 6 months for
inspection and preventive maintenance. Annual factory calibration checks are strongly recommended to
verify continued accuracy.
Note: If damage or wear is noted that may compromise the function of the instrument, do not use. Refer to RETURN
AUTHORIZATION AND REPLACEMENT INFORMATION section.

RECEIVING INSPECTION

After receiving the Zimmer Air Dermatome II, examine the unit for external signs of damage. Retain all packing material
until the contents have been verified and an initial operational check has been performed. If the device has been
damaged, forward an immediate request to the delivering carrier to perform an inspection and prepare a "Concealed
Damage" report. The terms of all Zimmer Surgical sales agreements concerning this unit are F.O.B., Dover, Ohio,
U.S.A., and the responsibility of the Zimmer Surgical ends with delivery to the first carrier. Thereafter, the shipment
becomes the property of the customer. Any claims for loss, damage, or non-delivery must be made against the
delivering carrier within 10 days of shipment.
Your local Zimmer sales representative may assist you in determining the cost of replacement or repair so that a proper
claim may be filed with the delivering carrier.
Note: Do not attempt to return damaged goods without first securing proper authorization from the delivering carrier.

SETUP INSTRUCTIONS

Prior to initial usage, the Zimmer Air Dermatome II must be cleaned and sterilized per instructions. (See
CLEANING AND DISINFECTION PROCESS INSTRUCTIONS FOR THE DERMATOME II and STERILIZATION
RECOMMENDATIONS.)
Observe sterile field precautions per hospital protocol.
Connect the handpiece to the hose by grasping both the swivel on the end of the handpiece and the hose firmly.
Push together and twist the hose to the right so that the pins are in the indentations (see figure 4).
When using a pressurized tank, the tank should be thoroughly wiped off with disinfectant and draped prior to
placement in the operating room. Always have the tank securely fastened to the wall, a stable cart or some other
stationary device to prevent falling and away from sources of heat.
Prior to setup in the operating room, open the tank valve very slowly and allow only enough gas to escape to
blow out any debris that may have accumulated in the valve. Stay clear of the opening, behind the tank, during
this procedure. Return the valve to the closed position. Install the regulator.
Once the regulator is securely installed, be certain the regulator is in the full OFF position. The tank valve may
then be opened slowly until it is fully open. This will allow for the gas to pressurize the regulator.
Never start a procedure if the pressure gauge indicates that there is less than 500 PSI (3450 kPa) in the tank.
Never allow the tank pressure to fall below 200 PSI (1380 kPa) or correct output pressure will not be maintained.
Insert the diffuser end of the hose into the quick-disconnect fitting on the regulator.
Activate the dermatome by completely depressing the throttle lever with the safety lock in the ON ( |-- ) position.
Pressure is established by gradually turning the regulator control knob. Always establish the designated pressure
using the pressure gauge with the instrument running. Recommended pressure is 100 PSI running, using a
recommended hose. When using an extension hose, increase the pressure at the source 1 PSI per foot
(22.6 kPa/m) of extension hose. For example, if a 10 foot (3.05 meter) extension hose is added to the standard
hose, indicated pressure at the source should be 110 PSI (759kPa) when the dermatome is running. Return the
safety lock to the SAFE position (O•), when not in use.
During setup procedure, visually inspect for damage and/or wear. If damage or wear that may compromise
the function of the instrument is noted, do not use.
3

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis